一、Industry Standards
Southeast Asian Countries Standard:
Vietnam :TCVN 4851-1989 "Water for Laboratory Analysis - Specifications and Test Methods".
Indonesia:SNI 6773:2008 "Drinking Water Standards"
European and American Countries Standard:
United States :ASTM D1193-2006 "Standard Specification for Reagent Water"."Preparation and Testing of Reagent Water for Medical Laboratories"
European Union :EN ISO 3696:1995 "Water for Analytical Laboratory Use - Specifications and Test Methods"ISO 3696:1995 ,European Pharmacopoeia (EP)
Middle Eastern Countries Standard:
Gulf Arab States :BH GSO ISO 3696:2024 "Water for Analytical Laboratory Use - Specifications and Test Methods"
African Countries Standard:South Africa :SANS 241
二、Basic Parameters
三、Advantages
System Control Description
Imported PLC automatic control system,A color touch screen enables intelligent online display of multiple parameters, including conductivity, resistivity, temperature, operating status, water tank liquid level and related notification messages, with built-in human-machine dialogue capability.
One-key disinfection function, no need to open the case or add chemicals. It realizes rapid disinfection of the RO membrane, circulation pipeline and water tank, inhibits the growth of bacteria and microorganisms, and ensures access to standard and high-efficiency water quality.
The system is furnished with RS232 and USB interfaces to connect with computers, enabling data record download and printing. It allows random access to data stored over the previous 5 years, which is in full compliance with Good Laboratory Practice (GLP) specifications.
Multi-level password control mechanism to prohibit unauthorized operations. This design is in full compliance with the implementation requirements of GLP management concepts and guarantees operational security.
Purity
The following features ensure that pure water is free from microbial contamination and meets the diverse water quality requirements of different customers: a built-in water tank equipped with an air filter and UV sterilizer, controllable delivery flow rate, consumable failure replacement reminders, modular ultrapure water columns, and a fast disinfection program.
SUS316L stainless steel pipeline material and stainless steel sterile water tank ensure the quality of the produced water and extend the overall service life of the equipment.
Flexibility
The modular pure water system connects multiple ultrapure water units into an integrated water treatment equipment network. Different water quality standards can be continuously adjusted according to application changes. Compared with traditional processes, the Vaughan Central Pure Water System not only saves valuable space but also ensures the purity of water quality standards.
Practicality
The system offers streamlined integration, effortless maintenance, and can be fully customized to meet laboratory-specific requirements.
Core Components
Genuine Dow membrane components imported from the US adopt a patented flow restriction technology design, keeping the reverse osmosis (RO) membrane submerged in water at all times. This prevents the RO membrane from being oxidized by exposure to air, effectively extending its service life by 30-50%, with a water recovery rate of up to 75%.
Genuine Italian-imported high-pressure pump automatically adjusts power based on pressure, ensuring stable water production of the RO membrane with low noise, energy efficiency and environmental friendliness.
High-quality EDI continuous electrodeionization modules to ensure continuous high-purity water output.
Technical Parameters
Model | Centra R | Centra E |
Product Water Flow (L/H): | 50/90/150/300/600/1000 | 50/90/150/300/500 |
Water Intake Flow Rate L/min) | ≤30 | ≤30 |
Power Supply | 380V/50Hz | 380V/50Hz |
Power Consumption | 0.3-2KW | 0.3-2KW |
Temperature Range | 5℃-45℃ | 5℃℃-459C |
Main Unit Dimensions (L×W×H | 1400*800*1900mm | 1800*800*1900mm |
Weight (kg) | 200-500 | 200-500 |
Safety Control | Power-off protection, internal water leakage alarm for the main unit, password key, low-voltage control circuit, system remote control operation, and pump overheat and overload protection. | |
System Detection | Circulating pipeline flow rate, circulation status, water tank storage capacity, product water quality, consumable replacement reminder, pressure display of each section. | |
Disinfection | One-button disinfection mode operation, automatic disinfection of external pipe network; the water tank adopts 360° rotating spray disinfection, with disinfection operation status display | |
Data Output | USB copy and download function | |
Inlet Water Indicators | ||
Water Source Requirements | 0.1-0.4Mpa, water temperature 5-45℃, total dissolved solids TDS≤200ppm, total organic carbon TOC≤500ppb | |
SDI Value | <5 | <5 |
Free Chlorine | <0.5PPM | <0.5PPM |
进水压力(Water Inlet Pressure) | 1.5-4bar | 1.5-4bar |
Price
Product Series | Model | Water Production Capacity | Price |
Centra R | Centra R90 | 90L/H | 174000.00 |
Centra R150 | 150L/H | 187200.00 | |
Centra R300 | 300L/H | 231600.00 | |
Centra R600 | 600L/H | 292800.00 | |
Centra R1000 | 1000L/H | 336000.00 | |
Centra E | Centra E90 | 90L/H | 218400.00 |
Centra E150 | 150L/H | 250800.00 | |
Centra E300 | 300L/H | 303600.00 | |
Centra E500 | 500L/H | 391200.00 |
Price:As it is a non-standard product, the price needs to be quoted based on specific requirements.